EDAN H100B User Manual - Package
of
99
H100B Pulse Oximeter User Manual
- 26 -
Checking the Ni-MH Rechargeable Battery
Package
The performance of a Ni-MH rechargeable battery
package may deteriorate. To check the performance of
the battery, follow the procedures below:
1. Disconnect the pulse oximeter from the patient
and stop all monitoring and measuring
procedures.
2. Place the pulse oximeter in the charger stand
and connect the AC mains. Allow the battery to
be charged uninterruptedly for above 2.5 hours.
For details about charging the Ni-MH
rechargeable battery package, please refer to
section 3.4.