EDAN H100B User Manual - Electromagnetic Immunity
of
99
H100B Pulse Oximeter User Manual
- 83 -
A2.3 Electromagnetic emissions-for
EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
Guidance and manufacturer’s declaration –
electromagnetic immunity
The oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of
the oximeter should assure that it is used in such an
environment.
Immunity
test
IEC/EN
60601
test
level
Compliance
level
Electromagnetic
environment -
guidance
Conducted
RF
IEC/EN
61000-4-6
Radiated
RF
IEC/EN
61000-4-3
3Vrms
150KHz
to
80MHz
3V/m
80 MHz
to
2.5GHz
3V
3V/m
Portable and
mobile RF
communications
equipment
should be used
no closer to any
part of the
oximeter,
including cables,