- 87 -
be considered. If the measured field strength in the location
in which the oximeter is used exceeds the applicable RF
compliance level above, the oximeter should be observed to
verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as
reorienting or relocating the oximeter.
A2.4 Recommended Separation Distances
Recommended separation distances between portable
and mobile RF communications equipment and the
oximeter
The oximeter is intended for use in an electromagnetic
environment in which radiated RF disturbances are
controlled. The customer or the user of the oximeter can
help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF
communications equipment (transmitters) and the oximeter
as recommended below, according to the maximum output
power of the communications equipment.
Separation distance according to
frequency of transmitter (m)