Setting up the selected probe
–
Press the ON / OFF key
=>
The
standard menu will appear on
the screen
In order to guarantee the maximum accuracy, the zero point of the probe should be set
from the mechanical to electrical zero before the first measurement is conducted. Use
the absolute mode setup to show the probe's raw value / the Absolute value.
!
–
Select menu
System with the u key
–
Select the submenu
Absolute mode
with the q key and press 4 to confirm
–
Set the probe mechanically to
"0"
–
Use the keys ut and t u to adjust
the scale
–
Press 4 to confirm
–
To exit the input menu, press the ON /
OFF key.