Mahr Millimar C 1200 Operating Instructions Manual - page 39
Other manuals for Millimar C 1200:
Operating Instructions Manual
Manual is about: Compact length measuring instrument
of
50
39
Mahr GmbH • Millimar C1200
3
6
3
q
p
Calibration Min
Nominal value
Calibration Max
-1.000
+0
1.000
Measurement
Display
Setup
System
Customer calibration
Please set the probe at
Zero
and accept
Measurement
Display
Setup
System
Absolute mode
Customer calibration
Factory calibration
Factory Settings
Info
6
3
q
p
ü
–
The connection of the jumper to the
Factory standard is as shown in the
illustration
=>
In the display appears:
Press set the probe at
Zero and
accept
– Press
4
to confirm
=>
The
Factory standard calibration
according to the customer's
specification is complete and
activated.
–
To exit the input menu, press the ON /
OFF key.