6.2
Customer calibration
6.2.1 Active customer calibration
–
Press the ON / OFF key
=>
The
standard menu will appear on
the screen
–
Press the u key to select the
System
menu
–
Select the submenu
Customer
calibration with the q key
– Press
4
to confirm
–
Select the submenu
Enable with the
q
key
– Press 4 to confirm or press Ó to
cancel selection
=> The
symbol 4 depicts the currently
active correction
–
To exit the input menu, press the ON /
OFF key.
If a customer specific calibration is required, this can be entered as described below.
The calibration can be conducted with the connected probe or alteratively with a factory
standard (e.g. Mahr 1283 WN Factory Calibration, Order no. 5312833).