Mahr Millimar C 1200 Operating Instructions Manual - page 36
36
Mahr GmbH • Millimar C1200
3
6
u
q
p
Calibration Min
Nominal value
Calibration Max
-1.000
+0
1.000
Measurement
Display
Setup
System
Customer calibration
+ 0 0
1
. 0 0 0
3
6
3
q
p
Calibration Min
Nominal value
Calibration Max
-1.000
+0
1.000
Measurement
Display
Setup
System
Customer calibration
Please set the probe at
Calibration
Min
and accept
3
6
3
q
p
Calibration Min
Nominal value
Calibration Max
-1.000
+0
1.000
Measurement
Display
Setup
System
Customer calibration
Please set the probe at
Calibration
Max
and accept
Measurement
Display
Setup
System
Absolute mode
Customer calibration
Factory calibration
Factory Settings
Info
6
3
q
p
ü
–
Enter the value for
Calibration Max
(e.g. +1.000 mm). Press the u key
to select the required position (digit).
Press the q p keys to enter the
required value.
– Press 4 to confirm or press Ó to
cancel selection
–
Move the
Probe to contact the Min
position (e.g. measuring plate)
=>
In the display appears:
Please set the probe at
Calibration
Min and accept
– Press
4
to confirm
–
Move the
Probe to contact the Max
position (e.g. 2.000 mm gage block)
=>
In the display appears:
Please set the probe at
Calibration
Max and accept
– Press
4
to confirm
=>
The
Probe calibration according
to the customer's specification is
complete and activated.
–
To exit the input menu, press the ON /
OFF key.