Mahr Millimar C 1200 Operating Instructions Manual - page 45
45
Mahr GmbH • Millimar C1200
Description
Command ('_' should be replaced with ' ' )
Command example ('_' should be replaced with ' ' )
Reply ('_' should be replaced with ' ' )
Permissable entry values
Notes
Request current digital value
?
±XXX.XX ±XXX.XX_< ±XXX.XX_= ±XXX.XX_>
-
-Reply according to current unit and resolution -If tolerance is enabled, the current state is given by the symbol following the value: = : in tolerance < : under tolerance > : over tolerance
Request device's model
ID?
-
C1200
-
-
Request
current
firmware
version
VER?
-
X.X
-
-
Set unit to µm
UM
-
UM
-
-
Set unit to mm
MM
-
MM
-
-
Set unit to inch
IN
-
IN
-
-
Set formula to +A
CHA+
-
CHA+
-
-
Set formula to -A
CHA-
-
CHA-
-
-
Turn device OFF
OFF
-
OFF
-
-
Set zero of
fset
RST
-
RST
-
-
Set factor
FAK_
FAK_2.5
FA
K
0.0001 to 10.0000 --> real number with maximum 4 decimal digits.
-
Request factor
FAK?
-
X.XXXX
-
-
Set master
MASTER_
MASTER_500
MASTER
In mm= -999.990 to +999.990 --> maximum 4 decimal digits In µm= -999990.0 to +999990.0 --> maximum 4 decimal digits In inch= -39.369685 to +39.369685 --> maximum 4 decimal digits
-Reply according to current unit and resolution
Request master
MASTER?
±XXX.XX
-
-Reply according to current unit and resolution
Set nominal value
NOMINAL_
NOMINAL_500
NOMINAL
In mm= -999.990 to +999.990 --> maximum 4 decimal digits In µm= -999990.0 to +999990.0 --> maximum 4 decimal digits In inch= -39.369685 to +39.369685 --> maximum 4 decimal digits
-Reply according to current unit and resolution
Request nominal value
NOMINAL?
-
±XXX.XX
-
-Reply according to current unit and resolution
Deactivate tolerance mode (OFF)
TOL0
-
TOL0
-
-
Activate tolerance mode (ON)
TOL1
-
TOL1
-
-
Configure
upper
and
lower
tolerance
limits
TOL_ ol>_ - werT ol> TOL_300_-300 TO L ol> or ol>: In mm= -5.0000 to +5.0000 --> maximum 4 decimal digits In µm= -5000.0 to +5000.0 --> maximum 1 decimal digit In inch= -0.196850 to +0.196850 --> maximum 6 decimal digits The upper tolerence limt value ol> must be larger than the lower tolerence limt value ol> Request tolerance limits TOL? - ±XXX.XX_±XXX.XX - -Reply according to current unit and resolution Factory settings FSET - - - -Instrument must be manually switched back on Request the serial number SN? - YYMMXXXX - - Deactivate the S/N output SN0 - SN0 - - Activate the S/N output SN1 - SN1 - - Set resolution RES_ RES_3 RES Resolution = {1, 2, 3} RES_1 RES_2 RES_3 mm 0.01 0.001 0.0001 µm 10 1 0.1 inch 0.0001 0.00001 0.000005 Request resolution RES? - 1 2 3 - - All commands and replies must end with 7.3 Data transmission parameters