Mahr Millimar C 1200 Operating Instructions Manual - page 37
37
Mahr GmbH • Millimar C1200
Measurement
Display
Setup
System
Customer calibration
Enable
Probe calibration
Factory standard calibration
Customer calibration Reset
6
3
q
p
Measurement
Display
Setup
System
Absolute mode
Customer calibration
Factory calibration
Factory Settings
Info
6
3
q
p
ü
3
6
u
q
p
Calibration Min
Nominal value
Calibration Max
-1.000
+0
1.000
Measurement
Display
Setup
System
Customer calibration
-
0 0 1 . 0 0 0
6.2.3 Conduct a Factory standard
calibration
–
Press the ON / OFF key
=>
The
standard menu will appear on
the screen
–
Press the u key to select the
System
menu
–
Select the submenu
Customer
calibration with the q key
– Press
4
to confirm
–
Press the q key to select the
Factory
calibration
– Press 4 to confirm or press Ó to
cancel selection
–
Enter the value for
Calibration Min
(e.g. - 1.000 mm). Press the u key
to select the required position (digit).
Press the q p keys to enter the
required value.
– Press 4 to confirm or press Ó to
cancel selection