Calibration Min
Nominal value
Calibration Max
-1.000
+0
1.000
Measurement
Display
Setup
System
Customer calibration
+
0 0 1 . 0 0 0
3
6
3
q
p
Calibration Min
Nominal value
Calibration Max
-1.000
+0
1.000
Measurement
Display
Setup
System
Customer calibration
Please set the probe at
Calibration
Min
and accept
3
6
3
q
p
Calibration Min
Nominal value
Calibration Max
-1.000
+0
1.000
Measurement
Display
Setup
System
Customer calibration
Please set the probe at
Calibration
Max
and accept
–
Enter the value for
Calibration Max
(e.g. +1.000 mm). Press the u key
to select the required position (digit).
Press the q p keys to enter the
required value.
– Press 4 to confirm or press Ó to
cancel selection
–
The connection of the jumper to the
Factory standard is as shown in the
illustration
=>
In the display appears:
Please set the probe at
Calibration
Min and accept
– Press
4
to confirm
–
The connection of the jumper to the
Factory standard is as shown in the
illustration
=>
In the display appears:
Please set the probe at
Calibration
Max and accept
– Press
4
to confirm