Summary of VEGABAR 82

  • Page 1

    Quick setup guide pressure transmitter with ceramic measuring cell vegabar 82 slave sensor for electronic differential pres- sure document id: 46332.

  • Page 2: Contents

    2 contents vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 contents 1 for your safety 1.1 authorised personnel ....................................................................................................... 3 1.2 appropriate use ...................................

  • Page 3: 1  For Your Safety

    3 1 for your safety vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 1 for your safety 1.1 authorised personnel all operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator. Duri...

  • Page 4

    4 1 for your safety vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 1.6 permissible process pressure the permissible process pressure is specified on the type label with "process pressure", see chapter "configuration". For safety reasons, this range may not be exceeded...

  • Page 5: 2  Product Description

    5 2 product description vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 2 product description 2.1 configuration the type label contains the most important data for identification and use of the instrument: 2 1 13 14 12 10 11 5 3 6 4 7 8 9 fig. 1: layout of the type lab...

  • Page 6: 3 Mounting

    6 3 mounting vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 3 mounting 3.1 general instructions for use of the instrument protect your instrument against moisture ingress through the following measures: • use the recommended cable (see chapter "connecting to power sup...

  • Page 7

    7 3 mounting vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 • configuration, master sensor suitable for electronic differential pressure • pressure type is identical for both sensors, i.E. Relative pressure/ relative pressure or absolute pressure/absolute pressure • m...

  • Page 8

    8 3 mounting vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 instrument selection nominal measuring range master: 1 bar nominal measuring range slave: 1 bar turn down: 1 bar/0.050 bar = 20 : 1 the measuring result (level, pressure difference) as well as measured value ...

  • Page 9

    9 4 connecting to power supply vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 4 connecting to power supply 4.1 connecting the connection to the master sensor is carried out through spring- loaded terminals in the respective housing. For this, use the supplied, confect...

  • Page 10

    10 4 connecting to power supply vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 7. Connect the screen to the internal ground terminal, connect the external ground terminal to potential equalisation 8. Tighten the compression nut of the cable entry gland. The seal ring ...

  • Page 11

    11 4 connecting to power supply vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 4.3 connection example 5 6 7 8 4 2 3 1 4 connect to master 5 1 2 + ( ) (-) 6 7 8 4...20ma fig. 5: connection example, electronic differential pressure 1 master sensor 2 slave sensor 3 conne...

  • Page 12: Module

    12 5 set up with the display and adjustment module vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 5 set up with the display and adjustment module 5.1 parameter adjustment - quick setup to quickly and easily adapt the sensor to the application, select the menu item "qu...

  • Page 13

    13 5 set up with the display and adjustment module vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 enter the percentage value and the corresponding pressure value for the max. Level. The quick setup for level measurement is finished. 4. Position correction in this menu...

  • Page 14

    14 5 set up with the display and adjustment module vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 the quick setup for differential pressure measurement is finished. 4. Unit, static pressure in this menu item, you determine the unit of the static, i.E. Superim- posed p...

  • Page 15

    15 5 set up with the display and adjustment module vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 the quick setup for interface measurement is finished. 4. Unit, static pressure in this menu item, you determine the unit of the static, i.E. Superim- posed pressure. 5. ...

  • Page 16

    16 5 set up with the display and adjustment module vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 display: settings, e.G., for language, measured value display, lighting diagnosis: information, e.G. On instrument status, pointer, measure- ment reliability, simulation ...

  • Page 17

    17 5 set up with the display and adjustment module vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 menu item parameter default setting damping integration time 0.0 s linearization linear current output current output - mode output characteristics 4 … 20 ma reaction whe...

  • Page 18

    18 5 set up with the display and adjustment module vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 menu item parameter default setting current out- put current output - meas. Vari- able lin. Percent - level current output - adjustment 0 … 100 % correspond to 4 … 20 ma ...

  • Page 19: 6 Supplement

    19 6 supplement vegabar 82 • slave sensor for electronic differential pressure 46332-en-160212 6 supplement 6.1 technical data electromechanical data - version ip 66/ip 67 options of the cable entry Ʋ cable entry m20 x 1.5, ½ npt Ʋ cable gland m20 x 1,5; ½ npt (cable ø see below table) Ʋ blind plug ...

  • Page 20

    Printing date: vega grieshaber kg am hohenstein 113 77761 schiltach germany 46332-en-160212 all statements concerning scope of delivery, application, practical use and operat- ing conditions of the sensors and processing systems correspond to the information available at the time of printing. Subjec...