T&D MCR-4V User Manual - Group Operations Via Master
23
Using MCR-4V
Group Operations via Master
Group operations such as the starting and stopping recording for the
connected units can only be performed via the Master. Upon the start of
a group recording, the main recording settings will be temporarily
copied to all connected Slaves. Upon the stopping of recording, all
copied settings will be returned to their original state.
Using “MCR for Windows”, it is possible to start
and stop recording by group by connecting the
Master to a PC via USB.
Slaves
Master
Settings Controlled by the Master:
Settings for Recording Mode, Recording Method, 50-60Hz Filter,
Recording Interval, and Recording Channels
Settings not Controlled by the Master:
Settings for Channel Name, Measurement Range, Scale Conversion
and Warning Monitoring
Other Settings:
Preheat and Recording Stop Trigger are only enabled on the Master.
IMPORTANT
- Operations related to the SD memory card (such as Data Export and Quick Format)
cannot be performed by group via the Master.
- To change settings for items that are not controlled by the Master, either change on the
target connected Slave unit, or use “MCR for Windows” by connecting each Slave unit
to the PC with a USB cable.
- Among the above “Settings not controlled by the Master”, the Channel Name and
Scale Conversion can only be set or changed via “MCR for Windows”.