LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A
Within the junction box, as shown in the figure, the primary lines and the
secondary lines must be laid at a distance and secured with cable ties in such
a way that contact between them is excluded.
Connections to the terminal strip are made as follows:
Terminal Cable wire color
Function
1
White
DC 24 V
3 Brown
GND
Laying lines safely
Junction box without
power supply:
Connecting the
DC 24 V power supply