ABB LS4000 Operating Instructions Manual - page 82
Other manuals for LS4000:
Operating Instructions Manual
Manual is about: Diode laser analyzers
of
212
82
LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A
Leading cables through cable glands
No. Meaning
1
Union nut
2
Connection socket
3
Braided shield
4
Springs
Proceed as follows:
Step
Procedure
1
Feed the cable through the union nut and through the connection
socket into the junction box until the springs come into contact
with the bare or exposed braided shield.
2
Screw the union nut onto the connection socket.
Tightening torques:
M20 power supply: 10 Nm
M20 T/P probes and analog/digital outputs: 12 Nm
M25 connecting cable transmitter/receiver unit: 12 Nm
3
Close the junction box.
Ensure that all of the unused cable glands are tightly sealed.
Metal cable glands