LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A
Installation – Installation flanges menu
Enter physical properties of the flanges
Name
Meaning/function
Log out
Log out and return to the main menu
Total flange path length
Enter the length of the measuring section through the
flanges and, if applicable, through the validation cells
or insertion tubes
Concentration in flanges
Enter the purging gas concentration in the flanges
(proportion of the measurement component in vol.-
%)
When using compressed air: Enter 20.9 %,
when using nitrogen: Enter 0 %.
Pressure input
Define the source of the value for the pressure
FIXED: Use a manually entered value
PROCESS: Use the process value
AMBIENT: Use the ambient value measured by the
analyzer
Temperature input
Define the source of the value for the temperature
FIXED: Use a manually entered value
PROCESS: Use the process value
AMBIENT: Use the ambient value measured by the
analyzer
Fixed pressure level
Enter a fixed value for the process gas pressure in the
flanges
Fixed temperature level
Enter a fixed value for the purging gas temperature in
the flanges
Total flange path length
+
Intended use
Description