Specifications
Operating voltage
AC 100–240 V ± 10 %; 50 to 60 Hz
Power consumption
30 VA
Protection class
I
Overvoltage category
II
Degree of pollution
2
Safe isolation
Safety extra-low voltage (SELV) on the low
voltage side
Overload protection
Voltage and current limitation
Specifications
Analog outputs
Three 4–20 mA outputs (one per measur-
ing component and for transmission), load
max. 500 Ω, not insulated
Analog inputs
Two 4–20 mA inputs for dynamic process
temperature and pressure correction, load
max. 100 Ω, not insulated
Digital outputs
Two outputs: 2 pin with N.O. contacts
30 V/1 A DC/AC; wired in accordance with
the requirements for class 2 circuits
1)
Service port
Ethernet 10BaseT
1) Class 2 circuits are energy-limited circuits with a maximum voltage of
AC 30 V or 42 V, a maximum current of 5 A and a maximum power of
100 VA.
Safety in accordance with Euro-
pean standards – CE
Tested to EN 61010-1:2010
EMC: Interference immunity
Tested to EN 61326-1:2013
Inspection level: Industrial area, fulfills at
least the evaluation criteria according to
Table 2 of EN 61326-1.
EMC: Emission interference
Tested to EN 61326-1:2013
Limit class B for interference field
strength and interference voltages is met.
The permissible environmental conditions for the transmitter unit
and receiver unit may differ from those of the junction box. In such
instances, compliance with the limit values of all modules must be
guaranteed by means of a suitable spatial arrangement on site.
If the junction box is set up outdoors, measures must be taken
where necessary to ensure the intended operation. This may in-
clude rain canopies, for example, or housings offering an adequate
degree of protection if necessary.
Power supply (in the
junction box)
Inputs and outputs
(in the junction box)
Safety and EMC