LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A
Adjusting the configuration
The calibration is performed using the instrument software.
Proceed as follows:
Step
Procedure
1
Start up the PC.
2
Open a web browser.
3
Enter the IP address of the analyzer.
The connection to the instrument software is now established.
The "Main menu" is displayed.
Proceed as follows:
Step
Procedure
1
Enter the operator password. The operator password is "xs2ls".
2
Click on "Log in".
You are now logged in as an operator.
The "Main menu" is displayed.
For the calibration process, you must temporarily change the
configuration of the analyzer.
Once the calibration process is completed, the original configura-
tion must be restored.
In order to facilitate restoration of the original configuration, choose one of
the following methods:
Make a note of the original settings before each change.
Before each change, save a screenshot of the original settings.
Proceed as follows:
Step
Procedure
1
Click on "Installation".
The "Installation procedure" menu is displayed.
2
Click on "Next" to start the installation wizard.
The "Installation – Process Parameters" menu is displayed.
Preliminary remarks
Open the instrument
software
Logging in
Adjusting the
configuration