ABB LS4000 Operating Instructions Manual - page 136
136
LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A
Connecting the power supply
This section only describes how to wire the power supply.
The device immediately starts up when the supply voltage is ap-
plied, but it is not possible to recognize when this has happened.
For this reason, the supply voltage must not yet be applied!
Note the cable specification
Infrared laser beam invisible to the human eye.
Risk of eye injury in the event of accidental startup.
Keep device disconnected until final inspection.
Within the junction box, as shown in the figure, the lines must be secured with
cable ties. (Remark: The figure shows a slightly different variant of the junction
box.)
Terminal Cable wire color
Function
1
White
DC 24 V
3 Brown
GND
Laying lines safely
Junction box without
power supply:
Connecting the
DC 24 V power supply