Determining cable runs and line runs
The cable runs must meet the following requirements:
No crossing of walkways
No risk of mechanical stress
No chemical or corrosive influences
No extreme temperature effects
Possibility of secure cable attachment
If there is a danger of explosion due to dust, precautionary measures must
be taken to prevent the development of static charges on the surface of
cables and leads.
Lay out all cables and lines for connecting the devices.
Select the cable lengths and line lengths according to the conditions in the
area.
For details of cable specifications, see Observing cable specifications on page
79.
The cable clips and line brackets must meet the following conditions:
They must allow the cables and lines to be laid securely.
It must be possible to open the cable clips and line brackets to allow the
cables and lines to be removed temporarily for calibration of a separate
calibration set or for servicing.
Determining cable
runs and line runs
Laying out cables and
lines
Requirements for
cable clips and line
brackets