ABB LS4000 Operating Instructions Manual - page 99
LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A
99
Connecting the electrical leads
Topic Page
Ensuring safety ..................................................................................................... 99
Establishing a protective grounding ................................................................ 99
Connecting the transmitter unit to the junction box .................................. 100
Connecting the receiver unit to the junction box ......................................... 101
Option: Connecting the T/P probes to the junction box............................ 102
Connecting the analog and digital outputs to the junction box .............. 103
Connecting the potential equalization terminal .......................................... 104
Connecting the power supply .......................................................................... 105
Ensuring safety
No explosion protection is present when performing this work!
Risk of explosion.
Make sure that no explosive gases or dusts are present when
performing this work.
Establishing a protective grounding
The protective grounding is established via the PE conductor of
the power cable.
Proceed as follows:
Step
Procedure
1
Connect the PE conductor of the power cable to the PE terminal of
the terminal strip.
Ensuring safety
Establishing a
protective grounding