LabStyle Dario User Manual - page 23
23
ago.
Keep the control solution away from children.
Shake the control solution well before using.
Wipe the container tip clean after use and be sure to close the container tightly
after each use.
No reconstitution or dilution of the control solution is necessary.
CONTROL SOLUTION TEST PRINCIPLE
Dario™ Glucose Control Solution contains a known amount of glucose that
reacts with chemicals on the Dario™ test strip.
When applying a drop of control solution to the test strip, following the same
procedure that is used for whole blood, the glucose in the control solution reacts
with chemicals in the test strip.
The result displayed on the meter should fall within the expected range printed on
the test strip cartridge label. Obtaining a single result within the expected
Range is
sufficient to verify system performance. A result outside the range printed
on the
Test Strip cartridge label may indicate the system is not performing correctly.
This control solution is not for use as a cleaning solution.
CHEMICAL COMPOSITION
Dario™ Glucose Control Solution is an aqueous solution, which contains D-glucose
as the reactive ingredient in the approximate concentration.
Buffer 2.1%
Polyvinylpyrrolidone 5%
Polyviny alcohol 5%
SDS 0.1%
Preservative 0.1%