LabStyle Dario User Manual - page 29
29
5. Shake the control solution container and apply a drop of control solution to a
piece of paper.
6. Hold the paper and move it so that the solution touches the tip of the yellow
window of the strip.
7. Important: Verify that enough control solution is applied to fill the window so
that it turns red. Do not put control solution on top of the test strip.
When the meter detects that the solution was applied, the “Measuring...”
message
appears. The orange rotating in the circle indicates that the test is in
progress.
8. The test result appears on the screen after six (6) seconds. Read your result and
look at the test strip.
The glucose test result is automatically saved to memory. The result appears on
the
screen and is automatically saved to memory.
It is possible to delete this control test from memory, if desired.
For more information, refer to the Dario™ Application User Guide at
www.mydario.com
.
9. When the test is done, remove the test strip.
10. Compare the results on the meter with the values displayed on the test strip
cartridge.
11. If the results are:
Within the acceptable range, then your meter is ready for use.
Outside of the acceptable range, perform the test again. If the problem
recurs,
contact customer support at
www.mydario.com
.