CHAPTER 8 – TECHNICAL INFORMATION & SPECIFICATIONS
This section provides technical information on your Dario™ System.
HOW THE GLUCOSE METER WORKS
The Dario™ system determines the amount of glucose in fresh capillary blood using
electrochemical biosensor technology. An electrical current is generated by
the
reaction of glucose with the reagent on the test strip (glucose oxidase). The
strength
of the current produced by the reaction depends on the amount of glucose
in the
blood sample. The glucose meter detects the current and converts it into a
blood
glucose reading.
COMPATIBILITY
Dario™ is compatible for use on smart mobile devices. For full list of compatible
devices please refer to
www.mydario.com/support/getting-started/
TECHNICAL SPECIFICATIONS
Meter Specifications
Result Range
10-600 mg/dL (0.5-33.3 mmol/L)
Calibration
Plasma-equivalent
Sample
Fresh capillary whole blood
Test Sample
0.3 microliter
Test Time
6 seconds
Assay Method
Glucose Oxidase Biosensor
Hematocrit Range
20% - 60%
Power Supply
From the smart phone or mobile device (Applicable for P/N
1021-02) / 3V CR1220 Lithium Battery (Applicable for P/N 1021-
03)
Glucose Units
mg/dL / mmol/L
Dimensions
104x211x11 mm (4.1x8.3x.4 in)
Weight
40 gram (1.4 oz)
Memory
Unlimited results
Memory is limited to smart mobile device memory
Average
7-day, 14-day or 30-day averaging
Display
Dario™ application on a smart mobile device screen
Function
Hypo warning: Set by user.