LabStyle Dario User Manual - page 51
51
Message Content
Possible Cause / Recommended Action
To reconnect your Dario. If the problem
recurs, some adjustments might be
required. Check the following settings on
your device:
Smart device volume setting
inappropriately.
Follow the instructions on the app
to adjust the settings in your
mobile
device:
1. Make sure that Mono Audio is set to
off:
· Verify that Mono Audio is set to off.
2. Adjust the audio balance:
· The audio volume balance must be set
in the middle between L and R.
3. Make sure that your Ringer and Alerts
volume is set to the maximum:
4. Allow Dario™ to use your microphone:
5. Make sure that your music volume is
set to the maximum.
Insufficient blood detected, try again with
a new strip
Not enough blood on the strip.
Repeat the measurement with a new strip.
The Strip is used, please insert a new
strip
A strip was used more than once.
Perform a new measurement with a new
strip
High! Measure with a new strip, if recurs
contact your physician Immediately
High glucose level.
Measurement result is above 600 mg/dL
(33.3 mmol/L). Repeat the measurement
with a new strip. If this recurs, contact
your physician Immediately. If your
measurement results are inconsistent
with the way you feel, contact customer
support at
www.mydario.com
.
Temperature out of range. Move to an area
between 10-45°C (50°-113°F), wait 30
minutes and try again
The meter was left in an extreme
temperature environment.
Move to a cooler or warmer area, wait
at least 30 minutes and repeat the
measurement with a new strip.