LabStyle Dario User Manual - page 30
30
12. Disconnect the meter from your smart mobile device.
13. Wipe the tip of the control solution container clean and tightly seal the container.
The Dario™ Glucose Control Solution ranges printed on the test strip
cartridge label, are intended for the Dario™ test strip cartridge only.
The glucose concentrations in these solutions, when measured by
systems other than Dario™ Glucose Monitoring System, may differ.
Applying too much or too little control solution to the test strip may
cause out- of-range results.
Do not use your Dario™ Glucose Monitoring System to test your blood if
your control solution result falls outside of the acceptable range.
14. Place the meter back into the housing:
a. Hold the meter at an angle with the part that connects to the smart
mobile device facing away from the ridged orange panel.
b. Push the meter into the housing at an angle and then push downwards until the
meter clicks into place.