EDAN iSE-301 User Manual - page 31
SE-301 Series Electrocardiograph User Manual Managing ECG Records
- 25 -
Chapter 5 Managing ECG Records
If you want to save the ECG records, you should set the Auto Save to To ECG or Ext. Memory.
The default value is To ECG. Then the ECG records will be saved in the File Manager or in the
external memory automatically.
Click File on the main screen to enter the file manager screen.
The File Manager allows records to be stored, deleted, printed and transmitted. When there is no
space for more records to be stored in the File Manager, the message MemFull will be displayed.
5.1 Transmitting ECG Records to the PC
NOTE: To transmit ECG records to the PC, data management software (DMS) of EDAN
must be installed in the PC. You should log into the Smart ECG Viewer software
before the transmission.
5.1.1 Transmitting ECG Records through the Network
Connect the PC to the electrocardiograph with an Ethernet cable recommended by the
manufacturer.
Auto Transmission:
1. Choose Setup > Transmission to enter the Transmission Setup window.
2. Set Auto Transmission to On and set Transmission mode to Ethernet.
3. Set the Server IP to the IP of the DMS.
4. Set the first three numbers of the Local IP to the first three numbers of the IP of Smart
ECG Viewer. The last number of the Local IP item can be set at random, but it can’t be the