EDAN iSE-301 User Manual - 6.3 Record Info Setup
SE-301 Series Electrocardiograph User Manual Settings
- 30 -
Lowpass Filter
The cutoff frequency can be set to 75Hz, 100Hz, 150Hz, 270Hz or 300Hz
Lowpass Filterrestricts the bandwidth of input signals.
All the input signals whose frequency is higher than the set cutoff frequency
will be attenuated.
NOTE: Only when EMG Filter is set to Off, can the setting of
Lowpass Filter be effective.
NOTE: To pass the distortion test, the electrocardiograph has to be configured with the
highest bandwidth in filter settings. Otherwise, ECG signal may be distorted.
6.3 Record Info Setup
6.3.1 Setup 1
Item
Description
Print Out
Choose from: On, Off
Select Off to disable the print function in the AUTO or RHYT mode.
Speed
Choose from: 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s and
50mm/s
NOTE: The speed is corresponding with the work mode.
Gain
Choose from: 1.25mm/mV, 2.5mm/mV, 5mm/mV, 10mm/mV,
20mm/mV, 10/5mm/mV and 20/10mm/mV
10/5mm/mV means that the gain of limb leads is set to 10mm/mV, while
the gain of chest leads is set to 5mm/mV.
Auto Record Style Choose from: 3×4, 3×4+1R, 1×12, 1×12+1R and 3×2+2×3
Auto Record
Sequence
Choose from: Sequential or Synchronous
Select Sequential, the lead group is printed one by one in a certain
sequence. The start time of a lead group is just the end time of the
previous lead group.
Select Synchronous, the lead group is printed one by one in a certain
sequence. All leads are printed with the same start time.
AGC
AGC means auto gain control.
Choose from: On or Off
Select On,the gain can be automatically adjusted according to actual
signals.