SE-301 Series Electrocardiograph User Manual Troubleshooting
- 40 -
A5: Please check whether there is any error information displayed on the screen.
If the hint Paper? or PaperErr is shown on the screen, please deal with it according to
the above-mentioned measures.
If the hint Transfer is shown on the screen, which means that the ECG is transmitting
the data to the PC, please wait a few seconds. You can start the printing after the data
has been transmitted.
If the problem still exists, please contact the manufacturer or the local distributor for
further disposal.
3) Transmitting Problems
Q1: The ECG doesn’t respond to any keys after a long time of transmitting. It transmits
nothing for there is no new data appearing on the interface of the PC software. What
should I do?
A1: Some error may occur during the transmission course, for example, the connection
between the ECG and the net cable may loosen. In this case, please restart the ECG. If
it doesn’t work, please restart the PC.
If the problem still exists, please contact the manufacturer or the local distributor for
further disposal.
4) Main Unit Problems
Q1: I was doing the examination when the machine suddenly gave out a sound and
displayed the hint Lead Off. What should I do?
A1: The leads are not connected well. Please check whether the electrodes are connected
to the patient skin well, and then make sure that the patient cable socket is connected
to the patient cable firmly.
If none of the above-mentioned measures take effect, please contact the manufacturer
or the local distributor for further disposal.
Q2: The touch screen is not sensitive after restoring to factory defaults. What should I do?
A2: Hold down the PRINT/STOP key while switching on the electrocardiograph, the
system will enter the touch screen calibration screen. Operate as indicated on the
screen.