EDAN iSE-301 User Manual - 6.5 Transmission Setup
SE-301 Series Electrocardiograph User Manual Settings
- 34 -
6.5 Transmission Setup
NOTE:
1. To transmit ECG data to the PC, the Smart ECG Viewer software produced of EDAN
must be installed in the PC. You should log into the Smart ECG Viewer software
before transmission.
2. If the power supply suddenly breakdown during data storage or transmission, file
system error may occur. In this case, the file system should be formatted.
Item
Description
Device No.
Enter Device No., it should be within 7 ASCII characters.
Auto Transmission
Choose from: On or Off
Select On, ECG data will be transmitted automatically after an ECG
report is printed out in the AUTO or RHYT mode.
Transmission Mode
Choose from: Ethernet or Wireless
FTP Information
Enter data in the FTP Path, FTP User Name textboxes.
IP Addresses
Set Server IP, Local IP, Set Gateway, Set Subnet Mask
For details, please refer to Section 5.1: "Transmitting ECG Records to
the PC".
6.6 Lead Setup
Item
Description
Rhythm
Lead1/2/3
Choose from: І, П, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, or V6,the default
NOTE: Rhythm Lead 1/2/3 must be different from each other.
Lead
Sequence
Choose from: Standard or Cabrera
Lead
Sequence
Lead group
1
Lead group
2
Lead group
3
Lead group
4
Standard
І, II, III
aVR, aVL,
aVF
V1, V2, V3
V4, V5, V6
Cabrera
aVL, І, -aVR
II, aVF, III
V1, V2, V3
V4, V5, V6
Lead off hint When it is set to On andlead off waves are detected in the presampled waves, a
lead off hint message will be displayed.