EDAN iSE-301 User Manual - Chapter 6 Settings
SE-301 Series Electrocardiograph User Manual Settings
- 29 -
Chapter 6 Settings
Click Setup on the main screen to display the System Setup screen.
NOTE: The underlined values are system default values.
6.1 Work Mode
Item
Description
Sampling Mode
(Only available in
the AUTO mode)
Choose from: Pre-Sample and Real-time
Select Pre-Sample, 10s ECG data sampled before pressing the
PRINT/STOP key will be printed out.
NOTE: When Sampling Mode is set to Pre-Sample, if you press
the PRINT/STOP key before the electrocardiograph
samples for 10s, the recorder will not respond.
Auto Arrhythmia
Detection
When enabled, if arrhythmia is detected in the AUTO mode, a hint will
pop up to ask you whether to print an extra rhythm report after the 12-lead
ECG report.
6.2 Filter
Item
Description
AC Filter
Choose from: On or Off
AC filter is used to suppress interference of AC power supply.
NOTE: AC frequency can be set to 50Hz or 60Hz on the Advanced
Setup screen according to local mains supply specifications.
DFT Filter
Choose from: 0.01Hz, 0.05Hz, 0.32Hz, or 0.67Hz
DFT Filter greatly reduces the baseline fluctuations without affecting the
ECG signals. The purpose of this filter is to keep the ECG signals on the
baseline of the printout.
The set value is the low limit of the frequency range.
EMG Filter
The cutoff frequency can be set to 25Hz, 35Hz, 45Hz or Off
EMG Filter suppresses disturbance caused by strong muscle tremor.