Fig. 25 - SET PO2 AL
Set PO2 Alarm, information includes (Fig. 25):
> Graphic SEt with PO2, MAX, and NX icons
> PO2 value (ATA) flashing
• A (< 2 sec) to step upward through Set Points from 1.20 to
1.60 one at a time.
• M (< 2 sec) to step back through Set Points one at a time.
• S (< 2 sec) to save the setting and revert to Set A Lead-in.
• S (2 sec) to revert to Set DTR Alarm.
SET U MENU (UTILITIES)
Sequence >> Lead-in >> Wet >> Units >> DS* >> SS* >> Algo* >> CF*
>> Glo >> SR
*Items apply to NORM only.
Set Points remain as set until changed.
Set U Lead-in, information includes (Fig. 26):
> Graphics SEt and U
• A (< 2 sec) to step forward to Set T Lead-in.
• M (< 2 sec) to step back to Set A Lead-in.
• S (< 2 sec) to access Set Wet.
Fig. 26 - SET U LEAD-IN