• Connect the LP hose to the BCD inflator
COMPLETE EXPLORER TEST
Follow the on-screen pre-dive sequenc
e. See PART 4 Section 2. If
the unit
fails the negative or HP leak test and the fault is not easily recognizable.
Follow these steps:
• Remove the LCV hose from the Canister or restrict the hose to stop the
counterlung venting through the LCV
• Turn on the gas.
• Put the BOV into closed circuit mode.
• Inflate the counterlungs by blowing into the loop until it is full then put the
BOV in open circuit mode.
• Fully submerge the unit and look for leaks.
!
NOTE: The OPV situated under the Canister will vent some bub-
bles. THIS IS NORMAL. Look for other leaks.
! WARNING: Complete all gear prep and pre-dive sequence checks
before diving.