Accu-Chek Mobile User Manual - page 183
Other manuals for Mobile:
User Manual, User Manual, Quick Reference Manual, Getting Started Manual, User Manual, User Manual, User Manual, Getting Started Manual
Manual is about: BLOOD GLUCOSE MONITORING SYSTEM
of
243
Cleaning the meter and
fi
nger pricker
183
The inside
You only need to clean the inside of the meter if
the area around the tip of the cassette is visibly
soiled,
the error message E-1 with the text “Cassette de-
fective Insert new cassette” is displayed; the error
message might be caused by a fault in advancing
the test area to the tip of the cassette,
the error message E-4 with the text “See Reference
Manual” is displayed; the error message might be
caused by
fl
uff, crumbs or the like inside the tip of
the cassette.
To enable you to clean the inside of the meter, you
have to open
Clean
in the
Tools
menu and then remove
the test cassette.
i
If the error message E-4 was displayed, a test
area is lost when you later reinsert the test
cassette.