Testing and storage conditions
209
15.3 Light conditions
Displayed text, numbers and symbols appear yellow.
Bright light shining on the display may make them
dif
fi
cult to read. Shield the meter if necessary, with
your body, for example.
Keep the meter away from very strong light
sources (e.g. direct sunlight, spot lights). These
may interfere with the proper functioning of the
meter and lead to error messages.
15.4 Local sources of interference
Strong electromagnetic
fi
elds may interfere with
the proper operation of the meter. Do not use
the meter close to sources of strong electro-
magnetic radiation.
To avoid electrostatic discharge, do not use the
meter in a very dry environment, particularly
one in which synthetic materials are present.