Accu-Chek Mobile User Manual - page 222
Other manuals for Mobile:
User Manual, User Manual, Quick Reference Manual, Getting Started Manual, User Manual, User Manual, User Manual, Getting Started Manual
Manual is about: BLOOD GLUCOSE MONITORING SYSTEM
of
243
Symbols, messages, problems and error messages
222
E-6
During the test, bright light, e.g. sunlight, fell on the
meter and has interfered with the test.
Move into the shade or shield the meter with your
body, for example. Repeat the test.
You applied blood or control solution to the test area
too soon, i.e. before the prompt
Apply blood
was dis-
played.
Repeat the test. Do not apply blood or control solution
to the test area until the prompt
Apply blood
is dis-
played.