Accu-Chek Mobile User Manual - page 215
Symbols, messages, problems and error messages
215
Temperature too low / too high:
This message ap-
pears at the start of a test if the ambient temperature
or the temperature of the meter is between +8 and
+10 °C or between +40 and +42 °C.
Use
, to select whether you wish to
Test
or
Quit
.
Test:
in this case, the meter advances a test area.
Quit:
in this case, the meter returns to the Main
menu. Close the tip cover.
If you decide to perform the test in spite of the warn-
ing, the result will be
fl
agged and saved with the sym-
bol
.
Do not use blood glucose results obtained de-
spite this warning as a basis for making thera-
peutic decisions. These results may be incorrect.
Incorrect results can cause the wrong therapeu-
tic decision to be taken and so produce serious
adverse health effects. Move to a place where
the temperature is between +10 and +40 °C and
wait for the temperature of the meter to adjust to
this temperature.