Hach ANATEL PAT700 User Manual - page 117
Page 115
Bottle Mode
•
TOC limit text box allows you to enter a limit an online TOC analysis must reach
before an excursion sample is captured. You may specify a value from 1 to 2000
ppb. The default setting is 500 ppb.
•
Run now icon immediately draws a water sample from the online source into the
excursion bottle and opens the bottle filling dialog box. See
.
Once the bottle has been filled, the excursion captured dialog box appears. See
•
Done icon returns you to the home screen. When an excursion condition occurs,
the analyzer performs an excursion sample capture. Following the excursion, the
analyzer always returns to the home screen.
•
Cancel icon takes you to the unload/replace bottles dialog box. All changes made