Hach ANATEL PAT700 User Manual - page 125
Page 123
Alarms
149
UnableToStartPlumbingTest
Analyzer cannot start plumbing test.
150
UnableToPrintDataLog
Analyzer cannot print data log.
151
LampExtinct
•
Analyzer detects that current lamp has failed.
•
Replace UV lamp.
153
Bottle1MarkedAsUsed
•
Bottle installed in location 1 is marked as used.
•
Replace bottle with a valid standard.
154
Bottle2MarkedAsUsed
•
Bottle installed in location 2 is marked as used.
•
Replace bottle with a valid standard.
155
Bottle3MarkedAsUsed
•
Bottle installed in location 3 is marked as used.
•
Replace bottle with a valid standard.
156
Bottle4MarkedAsUsed
•
Bottle installed in location 4 is marked as used.
•
Replace bottle with a valid standard.
157
ExcursionBottle1NotEmpty
•
Excursion bottle 1 is not empty.
•
Remove excursion bottle 1 and replace with empty bottle.
158
ExcursionBottle2NotEmpty
•
Excursion bottle 2 is not empty.
•
Remove excursion bottle 2 and replace with empty bottle.
159
ExcursionBottle3NotEmpty
•
Excursion bottle 3 is not empty.
•
Remove excursion bottle 3 and replace with empty bottle.
160
ExcursionBottle4NotEmpty
•
Excursion bottle 4 is not empty.
•
Remove excursion bottle 4 and replace with empty bottle.
172
RougeDetected
•
One or more of your 500 ppb standards took longer than
expected to oxidize.
•
Clean the cell.
165
LampSwitch
•
Analyzer has switched UV lamps because current lamp was
marginal or failed.
•
Replace UV lamp.
178
ExcursionCaptured
•
An excursion sample was captured in excursion bottle.
•
Remove sample from analyzer.
180
ConductivityMeterAlarm
Analyzer has not passed conductivity meter test.
181
BottleTestCancelled
Bottle test was cancelled by the user.
Table 16 Warning alarm messages
Code
Message
Alarm condition