Maintenance and Troubleshooting
10. Holding the lamp by its metal sleeve, insert the new lamp into the lamp holder hole
and clip into place. The lamp is self-aligning. Ensure that the lamp is locked in place
by the lamp retaining clip.
11. Reinsert the lamp module into the instrument by centering it over the manifold
assembly and gently sliding the module down to the manifold until it locks into place.