Hach ANATEL PAT700 User Manual - 8.4 Tests Tab
Page 128
Diagnostics
8.4 Tests tab
The tests dialog box allows you to test the following analyzer functions: RS–232 output,
digital I/O, 4-20 mA outputs, printer, plumbing systems, pump, and heat exchanger fan.
See
.
8.4.1 RS-232 test
The RS-232 test icon allows you to test serial communication from the serial port by
using the printer port as a return port. See
1.
Disconnect the printer and RS-232 serial connections and connect a serial cable
between the analyzer’s serial ports (from one serial port to the other).
2.
Press the start test icon.
Start test icon sends serial data out one serial port and reads the data on the other. If
the read data matches the sent data, the test passes.
Return icon returns you to the diagnostics dialog box.
Figure 104 Diagnostics dialog box, test tab
Figure 105 RS-232 test dialog box