Hach ANATEL PAT700 User Manual - 5.8 Security
Page 70
Setup
5.8 Security
The security icon allows you to modify the security settings for the system. The security
dialog box allows configuration of general security features, setting and modifying user
accounts, and viewing of the audit trail.
5.8.1 General tab
Enable security checkbox allows you to enable or disable the analyzer’s security
system. When security is enabled, you are required to sign on prior to making changes in
the analyzer, and only the administrator may access the security settings. When security
is disabled, you are not required to sign on. Disabling the analyzer’s security system
disables the audit trail and all controls on this dialog box except the enable data logging
checkbox. The default setting is disabled.
Enable data logging checkbox allows you to enable or disable data logging. When
enabled, all data (up to 5000 records) are automatically stored in the analyzer’s internal
data buffer. When disabled, no data are saved. The default setting is enabled.
Restore Factory Defaults button will return the analyzer configuration settings to the
factory defaults (see
PAT700 Default Settings, page 183
for a list of the default values).
When activated all valves are closed and the internal pump is turned off.
Printer icon prints the security settings.
Export icon opens the export settings dialog box. See
Return icon returns you to the setup dialog box and saves the changes. Any changes
are noted in the audit trail.
Figure 46 Security settings dialog box, general tab