Hach ANATEL PAT700 User Manual - Section 9
Page 139
Section 9
Data Review
9.1 Accessing the data review dialog box
1.
At the home screen, touch the << icon to open the sliding toolbar.
2.
Touch the data review icon.
Access to the data review dialog box may be delayed while available data is transferred
to the data log.
9.2 Data review dialog box
The data review dialog box is populated only when data logging is selected in the security
settings dialog box. The data log will hold up to 5,000 records. The data log contains
analysis results, alarms, bottle test results, and audit trail entries. Since the audit trail is
stored in the data log each audit trail entry uses one of the 5,000 data log entries. The
data log operates in a first in first out approach. Once the data log contains 5,000 entries
the oldest entry is replaced with the newest entry. See
.
Data review window lists all the data in the data log. The data appear as they would on
the printer. The data are filtered based on the settings in the filter dialog box. Data are
displayed in chronological order from earliest to most recent. The list scrolls to display the
earliest data.
Print icon sends the displayed data to the printer.
Export icon opens the export settings dialog box. See
.
Return icon returns you to the home screen.
Filter data icon opens the filter data dialog box. See
.
Figure 117 Data review dialog box