Hach ANATEL PAT700 User Manual - page 41
Page 39
Installation
Example:
Read command:
“HR”
Read format:
“HH:mm:SS
HH:mm:SS
u u u u s u u u u u u u u u u u u
HH:mm:SS
u u f u u u u u u f”
Response:
“00:01:00
00:00:00
50 0 1 1 Sensor 0 1 0 3 0 1 0 0 0 0 1 1
00:01:00
25 00 1.000 000 020 005 095 0 0 1.000”
Write format:
“HR HH:mm:SS
HH:mm:SS
u u u u s u u u u u u u u u u u u
HH:mm:SS
u u f u u u u u u f”
Write command:
“HR 00:01:00
00:00:00
50
0
1 1 Sensor 0 1 0 3 0 1 0 0 0 0 1 1
00:01:00
25 00 1.000 000 020 005 095 0 0 1.000”
Example:
Read command:
“HH”
Response:
“UVLamp1:
4 4300
1.000000
Good”
“UVLamp2:
0 4300
1.000000
Good”
Example:
Read command:
“HC”
Response:
“1.000”
Write format:
“HC f”
Write command:
“HC 0.95”
Example:
Read command:
“HT”
Response:
“1.000”
Write format:
“HT f”
Write command:
“HT 1.05”
“HH”
Get number of times UV lamp is turned on, hours left, voltage and status of both UV lamps
“HC”
Set/get user-calibrated
conductivity slope
Slope range should be between 0.9 and 1.1.
“HT”
Set/get user TOC slope
Slope range should be between 0.85 and 1.15.