Hach ANATEL PAT700 User Manual - page 94
Page 92
Bottle Mode
5.
To use default settings, touch the use default settings radio button in the TOC
calibration setup dialog box. To run any other configuration, touch the use custom
settings radio button, then define the desired configuration. (See
.)
The following procedure relies on the recommended default setting.
6.
Touch the forward icon. The OASIS door will unlock for five seconds. If the door is
not opened within five seconds, the unlock icon can be used to unlock the door
manually.
7.
Install the standards bottles following the instructions as shown in the load bottles
dialog box and ensuring that you place the standards bottles in the correct positions.
See
•
Bottle 1 = blank
•
Bottle 2 = 250 ppb sucrose
•
Bottle 3 = 500 ppb sucrose
•
Bottle 4 = 750 ppb sucrose
•
Bottle positions graphics show you the order in which to install the bottles.
•
Unlock icon energizes the bottle bay door lock for five seconds, allowing you to
open the door.
•
Back icon takes you to the run standards dialog box.
•
Forward icon takes you to the review test setup dialog box if RFID tagged bottles
are being used (see
), or to the enter bottle information dialog box if
bottles that are not RFID tagged are being used (see
•
Cancel icon returns you to the previous dialog box. All changes made in any of
the dialog boxes are lost.
8.
Close the bottle bay door.
9.
Touch the forward icon. If bottles have been properly loaded and the RFID tags can
be read, the review test setup dialog box appears. See
. If the bottles do not
have RFID tags, the enter bottle information dialog box appears. See
.
Figure 71 Load bottles dialog box, TOC calibration