Diagnostics
8.4.7 Fan test
The fan test icon allows you to test the analyzer’s fans. The analyzer contains an internal
online heat exchanger that activates to maintain proper water temperature to the analysis
cell. Two fans cool the heat exchanger.
Touching the fan test icon opens the fan test dialog box, where you can cycle the fans on
and off to ensure their proper operation. See
.
8.4.8 RFID test
The RFID test icon allows you to test the RFID reader system.
To conduct this test you will need to install four standards bottles equipped with RFID
tags. These bottle can be new or used standards bottles.
Open the OASIS system door and insert up to four standards bottles with RFID tags in
each bottle location. Touching the start button will initiate a test of the RFID system. The
system will automatically sense those locations where RFID tags are present. To verify
failure of the system the unlock button can be used to open the OASIS door during the
test. With the door open, and the bottles no longer positioned properly, the system will
return a failed test.
Figure 114 Fan test dialog box
Figure 115 RFID test dialog box