Hach ANATEL PAT700 User Manual - page 25
Page 23
Installation
2.5.1 Power supply wiring
WARNING
Potential property damage or electrical shock! Adhere to ground network
requirements for the facility.
illustrates power supply wiring terminals.
lists specifications for the power supply.
Select a wire style and gauge that meets local electrical codes. The power supply wiring
connector accepts 10 to 14 AWG wiring. Wire type can be stranded or solid. Strip 3/8”
(9.5 mm) of insulation from the ends of the wires and insert the stripped wire ends into
the appropriate connector terminals. Secure the wire with the integral screw and tug
firmly on each wire to make sure it is properly secured.
WARNING
Potential electrical shock! After power supply wiring has been installed, make sure
the field wiring box cover is in place and the screw is tightened with a screwdriver
to ensure that the cover is properly bonded to protective earth.
Figure 11 PAT700 power supply wiring terminals
Table 2 Power supply specifications
Description
Specification
Voltage input
100 to 230 VAC universal
Frequency
50 to 60 Hz
Power
65 W maximum, fused at 1.25 A
Line
Neutral
Ground