Modbus Protocol
Last TOC Analysis Parameters Block
The Last TOC Analysis Parameters block comprises of parameters that are updated
immediately after a successful TOC Analysis.
Address 300-303 (TOC) (Format: Double Float)
Get the instrument's Final TOC value. The Final TOC value consists of 4 words and read
as 64bit double float data.
Address 304-307 (Temperature) (Format: Double Float)
Get the instrument's Temperature value. The Temperature value consists of 4 words and
read as 64bit double float data.
Address 308-311 (Compensated Conductivity) (Format: Double Float)
Get the instrument's Compensated Conductivity value. The Compensated Conductivity
value consists of 4 words and read as 64bit double float data.
Address 312-315 (Un-Compensated Conductivity) (Format: Double Float)
Get the instrument's Un-Compensated Conductivity value. The Un-Compensated
Conductivity value consists of 4 words and read as 64bit double float data.
Address 316 (Curve Type) (Format: 16 Bit Word)
Get the instrument's Curve Type. The Curve Type value consists of 1 word.
Address 317-320 (TOC Analysis Timestamp) (Format: Date & Time)
Get the instrument's TOC Analysis Timestamp. The Timestamp value consists of 4
words.
Address 321-399 - Expansion
Reserved for future expansion.
Last TOC Analysis Parameters Block
Address
Register Description
Access
Size
(Words)
0300
Final TOC
R
4
0304
Temperature
R
4
0308
Compensated Conductivity
R
4
0312
Uncompensated Conductivity
R
4
0316
Curve type
R
1
0317
TOC Analysis Timestamp
R
4
0321-0399
Future Expansion