Hach ANATEL PAT700 User Manual - 5.7 Printer
Page 67
Setup
5.7 Printer
The printer icon allows you to configure the printer settings for the analyzer. You can
independently configure the printing of analysis results and the frequency of printing for
online TOC and conductivity modes.
Printer connection is made using the 9-pin D-subminiature serial port. The RS-232
interface communicates at 9600 baud, 8 data bits, 1 stop bit, no parity.
The printout report requires a 40-column printer.
•
If the printer is purchased from HACH Company, the switches are preset at the
factory.
•
If a Seiko Instruments DPU-414 Type II thermal printer is installed, the electronic DIP
switches must be set as listed in
Table 13 Printer DIP switch settings
DIP Switch 1
Switch
Setting
Function
1
Off
Serial input
2
On
High print speed
3
On
Auto loading enabled
4
Off
Auto linefeed disabled
5
On
Enabled setting
6
Off
Print density
7
On
Print density
8
On
Print density
DIP Switch 2
Switch
Setting
Function
1
On
40-column printing
2
On
Backup font enabled
3
On
Normal characters
4
Off
Zero = Slash
5
On
International character set = USA
6
On
International character set = USA
7
On
International character set = USA
8
Off
International character set = USA
DIP Switch 3
Switch
Setting
Function
1
On
8 data bits
2
Off
Parity setting = Yes
3
On
Parity condition = Odd
4
Off
Flow control (X on/X off)
5
Off
9600 baud rate
6
On
9600 baud rate
7
On
9600 baud rate
8
On
9600 baud rate