Page 41
Installation
Example:
Read command:
“PV”
Response:
2.8.7 Diagnostic commands
Example:
Read command:
“RM”
Read format:
“Mode State”
Response:
“0 0”
“PV”
Show TOC validation history
“RM”
Get analyzer mode and state
Mode
State
0 = Offline
0 = Idle
1 = Online TOC
1 = Start sample
2 = Online conductivity
2= Calibrate
3 = Digital TOC
3 = Start flushing
4 = Manual TOC
4 = Flushing
5 = Manual bottle
5 = Wait 15 seconds
6 = Manual clean
6 = Reading TOC
7 = Manual diagnostic
8 = Analysis done
8 = Standby
9 = Initialization error
10 = Error