Alarms
7.4.2 Critical error alarms
A critical error occurs any time the analyzer cannot function due to an error in the system.
The error appears as a full-screen, flashing message. All operation ceases until the error
condition has been corrected. The current analysis is aborted. Touching the screen
provides access to the alarm acknowledge dialog box. See
.
7.4.3 Warning alarms
A warning occurs any time one of the subsystems is not operating at a normal level.
When a warning condition is present, a warning indicator appears on the display.
Operation may continue with a warning condition present. When the warning condition
has been corrected, the indicator disappears. See
.
Table 15 Critical error alarm messages
Code
Message
Alarm condition
110
UnableToStartIoModule
Analyzer cannot start IO module.
111
UnableToStartTocModule
Analyzer cannot start TOC module
114
UnableToStartEthernetModule
Analyzer cannot start Ethernet module.
152
LampUnableToSwitch
Analyzer cannot switch from one lamp to the other.
166
UnableToInitializePump
Failed to initialize pump to correct speed.
173
RangeTable_FailedToLoad
RangeTable.txt file could not be loaded.
174
AnalyzerTemperature
Temperature inside analyzer is out of normal operating range.
175
BothLampsBad
•
Both UV lamps have failed.
•
Replace both lamps as soon as possible.
176
NoCondTempReadings
•
Analyzer cannot measure conductivity and temperature.
•
Restart analyzer.
•
If restarting doesn't resolve error, phone customer service at
1–800–866–8854.
179
HardwareInitializationFailure
An error occurred during hardware initialization.
Table 16 Warning alarm messages
Code
Message
Alarm condition
27
FALSE_CONDUCTIVITY
Measured conductivity is less than 0.
28
BAD_TEMP_CONVERT
Conductivity out of bounds for temperature conversion.
29
BAD_TOC_CONVERT
34
ANALYSIS_TIMEOUT
Unable to finish oxidation in 35 minutes.
35
LOW_OX_RATE
TOC too low (out of range).
36
CHEM_INTERFERE
•
Negative TOC (out of range).
•
Potential presence of organic acids.
37
TEMP_LOW
Temperature out of range low.
38
TEMP_HIGH
Temperature out of range high.
39
RESIST_LOW
Resistivity out of range low.
40
RESIST_HIGH
Resistivity out of range high.
100
IncorrectBottleLoadedinPosition1
•
Incorrect bottle detected in position 1.
•
Analyzer cannot perform bottle test operation.
101
IncorrectBottleLoadedinPosition2
•
Incorrect bottle detected in position 2.
•
Analyzer cannot perform bottle test operation.
102
IncorrectBottleLoadedinPosition3
•
Incorrect bottle detected in position 3.
•
Analyzer cannot perform bottle test operation.